Operation

5xx Pressure and flow errors

500

W

Pressure sensor not
connected,
(pressure p1 voltage
[V])

58

144

0.5V

Pirani P1 not connected, IF
board or MSB defective

520

W

Pressure too high,
(pressure p1 [mbar])

73

148

18
mbar

Pressure p1 too high

521

W

Pressure rise, anode
break down,
(anode voltage [V])

73

148

< Nom.
Value -
20V

Pressure p1 too high,
message after 1.4s

522

W

Pressure rise emission
break down,
(emission current [A])

73

148

< 90%
Nom.
Value
> 110%
Nom.
value

Emission was stable
previously, pressure p1 too
high, message after 5s

540

W

Pressure too low,
sniffer blocked,
(pressure p1 [mbar])

63

62

Param
eter
Sniffer
flow
warnin
g

Sniffer clogged, sniffer
valve defective

541

E

Sniffer blocked

62

146

Sniffer blocked, sniffer
valve defective (pressure
lower than half of the
configured warning value)

542

W

Sniffer broken,
(pressure p1 [mbar])

64

147

Sniffer broken

6xx Calibration errors

600

W

Calfac too low,
(calibration factor)

81

153

0.01

Calibration leak or machine
factor set incorrectly

601

W

Calfac too high,
(calibration factor)

81

153

5000

Calibration leak or machine
factor set incorrectly, split
flow factor too high

602

W

Calfac lower than last
calibration,
(calibration factor)

81

153

< 50%
old
value

Calibration leak, machine
factor or split flow factor
has changed

603

W

Calfac higher than last
calibration,
(calibration factor)

81

153

> 200%
old
value

Calibration leak, machine
factor or split flow factor
has changed

604

W

No int cal due to valve
control,
(valve control by user)

81

153

Test leak is not enabled

605

W

Testleak too small

Test leak defective or
signal too weak

610

W

Machine factor too low,
(machine factor)

81

153

1.00E-
04

Machine factor adjustment
inaccurate

Error

number

Warning/

Error

Error message
LDS3000

Error

number

LDS1000

protocol

Error

number

ASCII

protocol*

Limit
values

Cause