Operation
5xx Pressure and flow errors
500
W
Pressure sensor not
connected,
(pressure p1 voltage
[V])
58
144
0.5V
Pirani P1 not connected, IF
board or MSB defective
520
W
Pressure too high,
(pressure p1 [mbar])
73
148
18
mbar
Pressure p1 too high
521
W
Pressure rise, anode
break down,
(anode voltage [V])
73
148
< Nom.
Value -
20V
Pressure p1 too high,
message after 1.4s
522
W
Pressure rise emission
break down,
(emission current [A])
73
148
< 90%
Nom.
Value
> 110%
Nom.
value
Emission was stable
previously, pressure p1 too
high, message after 5s
540
W
Pressure too low,
sniffer blocked,
(pressure p1 [mbar])
63
62
Param
eter
Sniffer
flow
warnin
g
Sniffer clogged, sniffer
valve defective
541
E
Sniffer blocked
62
146
Sniffer blocked, sniffer
valve defective (pressure
lower than half of the
configured warning value)
542
W
Sniffer broken,
(pressure p1 [mbar])
64
147
Sniffer broken
6xx Calibration errors
600
W
Calfac too low,
(calibration factor)
81
153
0.01
Calibration leak or machine
factor set incorrectly
601
W
Calfac too high,
(calibration factor)
81
153
5000
Calibration leak or machine
factor set incorrectly, split
flow factor too high
602
W
Calfac lower than last
calibration,
(calibration factor)
81
153
< 50%
old
value
Calibration leak, machine
factor or split flow factor
has changed
603
W
Calfac higher than last
calibration,
(calibration factor)
81
153
> 200%
old
value
Calibration leak, machine
factor or split flow factor
has changed
604
W
No int cal due to valve
control,
(valve control by user)
81
153
Test leak is not enabled
605
W
Testleak too small
Test leak defective or
signal too weak
610
W
Machine factor too low,
(machine factor)
81
153
1.00E-
04
Machine factor adjustment
inaccurate
Error
number
Warning/
Error
Error message
LDS3000
Error
number
LDS1000
protocol
Error
number
ASCII
protocol*
Limit
values
Cause