Operation

6.15

Enable/disable correction of the leakage rate in Standby

Enabled = machine/sniff factor is considered in Standby.
Disable = machine/sniff factor is not considered in Standby.

► Enable or disable correction of the leakage rate in Standby as needed:

S

ETTINGS

> S

ET

UP

> O

PERATION

MODES

> LR

CORRECTION

> M

ACHINE

FACTOR

IN

STANDBY

6.16

Setting the leakage rate threshold value

The mass spectrometer module has four independent leakage rate threshold values.

► Set the leakage rate threshold value as needed:

S

ETTINGS

> T

RIGGER

> T

RIGGER

1/2/3/4 > T

RIGGER

1/2/3/4

LEVEL

6.17

Enable/disable Z

ERO

key (Sniffer key)

► Enable or disable Z

ERO

key on sniffer line as needed:

S

ETTINGS

> S

ET

UP

> O

PERATION

MODES

> S

NIFFER

> S

NIFFER

KEY

> Z

ERO

KEY

SNIFFER

6.18

Setting capillary surveillance

To detect clogged sniffer lines, the backing pressure is monitored in sniffer mode.
You can adjust the pressure threshold that will result in a warning if values fall below
it.

► Setting surveillance of the backing pressure:

S

ETTINGS

> S

ET

UP

> O

PERATION

MODES

> S

NIFFER

> C

APILLARY

>

P

RESSURE

CAPILLARY

BLOCKED

6.19

Compatibility with LDS1000 and LDS2010

Switching the compatibility mode will cause the unit to load the corresponding factory
settings of the LDS1000, LDS2010 or LDS3000.

6.20

Warning and malfunction messages

Error

number

Warning/

Error

Error message
LDS3000

Error

number

LDS1000

protocol

Error

number

ASCII

protocol*

Limit
values

Cause

1xx system error (RAM, ROM, EEPROM, clock, ...)

102

W

Timeout EEPROM
MSB,
(number of parameter)

84

43

EEPROM on IF board or
MSB defective

104

W

One EEPROM
parameter initialized,
(number of parameter)

84

43

Following software update
or EEPROM defective

106

W

EEPROM parameters
initialized,
(number of parameter)

84

43

Following software update
or EEPROM defective