Factory layout, Possible layouts, 1 factory layout – INFICON LDS3000 Modular Leak Detector User Manual
Page 22: 2 possible layouts

22
Operation
6.2.1
Factory layout
Analog output 1: Leak rate mantissa
Analog output 2: Leak rate exponent
6.2.2
Possible layouts
Off
The analog outputs are disabled (output voltage = 0 V).
Pressure p1 / Pressure p2
1 ... 10 V; 0.5 V / decade; 1 V = 1 x 10
-3
mbar
Leak rate mantissa
1 ... 10 V; linear; in used unit
Information Useful only if the other analog output is assigned “Leak rate exponent”.
Leak rate exponent
1 ... 10 V; 0.5 V / decade; step function; 1 V = 1 x 10
-12
; in used unit
Information Useful only if the other analog output is assigned “Leak rate mantissa”
or “Leak rate ma. hys.”.
Leak rate linear
x ... 10 V; linear; in used unit
The upper limit (= 10 V) is set via the parameter “Upper limit exponent”.
Leak rate log.
x ... 10 V; logarithmic; in used unit
The upper limit (= 10 V) and the scale (V / decades) are set via the parameters
“Upper limit exponent” and “Scale for leak rate”.
For example:
Upper limit set to 10
-5
mbar·l/s (= 10 V).
Scale set to 5 V / decade.
Lower limit is 10
-3
mbar·l/s (= 0 V).
- TGF10 Tracer Gas Filler (36 pages)
- Sensistor ILS500 F Leak Detection Filler (90 pages)
- T-Guard Leak Detection Sensor (85 pages)
- T-Guard Leak Detection Sensor Interface description (40 pages)
- Sensistor ISH2000P Hydrogen Leak Detector, Panel Model (51 pages)
- Sensistor ISH2000 HySpeed Hydrogen Leak Detector (54 pages)
- LDS3000 Modular Leak Detector Interface description (56 pages)
- BM1000 Bus module (14 pages)
- I/O1000 I/O module (18 pages)
- CU1000 Control unit (24 pages)
- Helium Leak Detector Modul1000 (130 pages)
- Helium Leak Detector Modul1000 Interface description (40 pages)
- UL5000 Dry Helium Leak Detector (108 pages)
- UL5000 Dry Helium Leak Detector Interface description (14 pages)
- UL1000 Fab Dry Helium Leak Detector (119 pages)
- HLD6000 Refrigerant Leak Detector (76 pages)
- HLD6000 Refrigerant Leak Detector Interface Description (40 pages)
- IO1000 I/O module (18 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector (92 pages)
- Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description (36 pages)
- Sensistor XRS9012 Hydrogen Leak Detector User Manual (28 pages)
- Sensistor XRS9012 Hydrogen Leak Detector Maintenance manual (14 pages)
- Extrima Ex-certified Hydrogen Leak Detector (62 pages)
- Sensistor ILS500 Leak Detection System (107 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (58 pages)
- Sensistor ISH2000 Hydrogen Leak Detector (108 pages)
- Sensistor Sentrac Hydrogen Leak Detector (86 pages)
- Protec P3000(XL) Helium Leak Detector (132 pages)
- Pilot Plus Vacuum Gauge (2 pages)
- CO Check Carbon Monoxide Meter (2 pages)
- GAS-Mate Combustible Gas Leak Detector (12 pages)
- Whisper Ultrasonic Leak Detector (8 pages)
- Vortex AC Refrigerant Recovery Machine 115V (20 pages)
- Vortex AC Refrigerant Recovery Machine 230V (16 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (2 pages)
- Wey-TEK Refrigerant Charging Scale & Optional Charging Module (44 pages)
- D-TEK CO2 Refrigerant Leak Detector (12 pages)
- TEK-Mate Refrigerant Leak Detector (12 pages)
- Compass Refrigerant Leak Detector (12 pages)
- D-TEK Select Refrigerant Leak Detector (12 pages)
- Explorer Portable Gas Chromatograph (369 pages)
- MicroFID II Portable Flame Ionization Detector (89 pages)
- DataFID Portable Flame Ionization Detector for Landfill Emissions Monitoring (91 pages)
- Hydrostik Hydrogen Fuel Cylinder Installation (7 pages)