Description

4.3.2

Electrical data

4.3.3

Physical data

4.3.4

Ambient conditions

Supply voltage

24 V ± 10% DC

Power input

max. 10 A

Noise level

< 60 dB(A)

Detectable gases

4

He, H

2

, Mass 3 (e. g. H-D,

3

He or H

3

)

Max. inlet pressure
(varying with the operation mode and
the speed of the turbo molecular
pump)

0.2 mbar ... 18 mbar

Operation in vacuum mode

Minimum detectable leak rate
(varying with the operation mode and
the speed of the turbo molecular
pump):

Helium

< 5 x 10

-12

mbar·l/s

Time until ready for operation

150 s

Operation in Sniffer mode

Minimum detectable leak rate:

Helium

< 1 x 10

-7

mbar·l/s

Response time in Sniffer mode

G

ROSS

: < 5 s, F

INE

/U

LTRA

: < 1 s

Permissible ambient temperature
(during operation)

10 °C ... 45 °C

Permissible storage temperature

-20 °C ... 60 °C

Max. relative humidity

at temperatures:

< +31 °C

80%

+31 °C to +40 °C

decreasing linearly from 80% ... 50%

> +40 °C

50%

Type of protection

IP 40

Pollution degree

II

Max. altitude above sea level

2000 m

Max. induction