11 select sample gas, 12 decontaminating backing pump from test gas, 13 selecting units for leakage rate – INFICON LDS3000 Modular Leak Detector User Manual
Page 35: 14 selecting units for pressure
Operation
35
6.11
Select sample gas
The mass spectrometer module is capable of measuring the following gases:
• Hydrogen (H
2
)
• Mass 3 (H-D,
3
He oder H
3
)
• Helium (
4
He)
► Select desired mass:
S
ETTINGS
> M
ASS
> M
ASS
The machine, calibration and sniff factor are dependent on the configured mass and
are saved in the mass spectrometer module.
6.12
Decontaminating backing pump from test gas
The mass spectrometer module can control an electric 24V gas ballast valve of the
backing pump via the O
UTPUT
connection.
► Controlling the gas ballast valve using digital outputs.
6.13
Selecting units for leakage rate
For control unit:
► Select desired unit of the leakage rate in vacuum mode:
V
IEW
SETTINGS
> U
NITS
(
DISPLAY
) > L
EAK
RATE
UNIT
VAC
► Select desired unit of the leakage rate in sniff mode:
V
IEW
SETTINGS
> U
NITS
(
DISPLAY
) > L
EAK
RATE
UNIT
SNIFF
For I/O or bus module:
► Select desired unit of the leakage rate in vacuum mode:
S
ETTINGS
> S
ET
UP
> I
NTERFACES
> U
NITS
(
INTERFACE
) > L
EAK
RATE
UNIT
V
AC
► Select desired unit of the leakage rate in sniff mode:
S
ETTINGS
> S
ET
UP
> I
NTERFACES
> U
NITS
(
INTERFACE
) > L
EAK
RATE
UNIT
S
NIFF
6.14
Selecting units for pressure
For control unit:
► Select desired unit for the pressure:
V
IEW
SETTINGS
> U
NITS
(
DISPLAY
) > P
RESSURE
UNIT
For I/O or bus module:
► Select desired unit for the pressure:
S
ETTINGS
> S
ET
UP
> I
NTERFACES
> U
NITS
(
INTERFACE
) > P
RESSURE
UNIT