Operation

35

6.11

Select sample gas

The mass spectrometer module is capable of measuring the following gases:

• Hydrogen (H

2

)

• Mass 3 (H-D,

3

He oder H

3

)

• Helium (

4

He)

► Select desired mass:

S

ETTINGS

> M

ASS

> M

ASS

The machine, calibration and sniff factor are dependent on the configured mass and
are saved in the mass spectrometer module.

6.12

Decontaminating backing pump from test gas

The mass spectrometer module can control an electric 24V gas ballast valve of the
backing pump via the O

UTPUT

connection.

► Controlling the gas ballast valve using digital outputs.

6.13

Selecting units for leakage rate

For control unit:

► Select desired unit of the leakage rate in vacuum mode:

V

IEW

SETTINGS

> U

NITS

(

DISPLAY

) > L

EAK

RATE

UNIT

VAC

► Select desired unit of the leakage rate in sniff mode:

V

IEW

SETTINGS

> U

NITS

(

DISPLAY

) > L

EAK

RATE

UNIT

SNIFF

For I/O or bus module:

► Select desired unit of the leakage rate in vacuum mode:

S

ETTINGS

> S

ET

UP

> I

NTERFACES

> U

NITS

(

INTERFACE

) > L

EAK

RATE

UNIT

V

AC

► Select desired unit of the leakage rate in sniff mode:

S

ETTINGS

> S

ET

UP

> I

NTERFACES

> U

NITS

(

INTERFACE

) > L

EAK

RATE

UNIT

S

NIFF

6.14

Selecting units for pressure

For control unit:

► Select desired unit for the pressure:

V

IEW

SETTINGS

> U

NITS

(

DISPLAY

) > P

RESSURE

UNIT

For I/O or bus module:

► Select desired unit for the pressure:

S

ETTINGS

> S

ET

UP

> I

NTERFACES

> U

NITS

(

INTERFACE

) > P

RESSURE

UNIT