Operation

33

6.10.1 Internal calibration

Internal test leak connected.

1

Enter leakage rate of the test leak via digital interface (control unit CU1000,
I/O module IO1000 or bus module BM1000).

2

Start calibration.

 Calibration is performed automatically.

6.10.2 External calibration

Vacuum mode: External test leak installed and open in or on the test system.

Sniffer mode: Sniffer line measures sniffer leak.

1

Enter leakage rate of the external test leak via digital interface (control
unit CU1000, I/O module IO1000 or bus module BM1000).

2

Wait until leakage rate signal is tuned and stable.

3

Start calibration.

 Request C

LOSE

TEST

LEAK

4

Vacuum mode: Close test leak inside the test system.

5

Sniffer mode: Remove sniffer line from sniffer leak.

 Leakage rate signal decreases.

6

Confirm measured background value is stable.

6.10.2.1 External dynamic calibration

Vacuum mode: External test leak installed and open in or on the test system.

Sniffer mode: Sniffer line measures sniffer leak.

1

Enter leakage rate of the external test leak via digital interface (control
unit CU1000, I/O module IO1000 or bus module BM1000).

2

Wait until leakage rate signal is tuned and stable.

3

If you wish to apply non-tuned leakage rate signal: Start calibration.

 This leakage rate signal will be used for the calibration.

 Request C

LOSE

TEST

LEAK

.

4

To determine the helium background, close the inlet valve on the test system.

 Leakage rate signal decreases.

5

Confirm significantly decreased leakage rate signal.