1 internal calibration 6.10.2 external calibration, 2 external calibration – INFICON LDS3000 Modular Leak Detector User Manual
Page 33
Operation
33
6.10.1 Internal calibration
Internal test leak connected.
1
Enter leakage rate of the test leak via digital interface (control unit CU1000,
I/O module IO1000 or bus module BM1000).
2
Start calibration.
Calibration is performed automatically.
6.10.2 External calibration
Vacuum mode: External test leak installed and open in or on the test system.
Sniffer mode: Sniffer line measures sniffer leak.
1
Enter leakage rate of the external test leak via digital interface (control
unit CU1000, I/O module IO1000 or bus module BM1000).
2
Wait until leakage rate signal is tuned and stable.
3
Start calibration.
Request C
LOSE
TEST
LEAK
4
Vacuum mode: Close test leak inside the test system.
5
Sniffer mode: Remove sniffer line from sniffer leak.
Leakage rate signal decreases.
6
Confirm measured background value is stable.
6.10.2.1 External dynamic calibration
Vacuum mode: External test leak installed and open in or on the test system.
Sniffer mode: Sniffer line measures sniffer leak.
1
Enter leakage rate of the external test leak via digital interface (control
unit CU1000, I/O module IO1000 or bus module BM1000).
2
Wait until leakage rate signal is tuned and stable.
3
If you wish to apply non-tuned leakage rate signal: Start calibration.
This leakage rate signal will be used for the calibration.
Request C
LOSE
TEST
LEAK
.
4
To determine the helium background, close the inlet valve on the test system.
Leakage rate signal decreases.
5
Confirm significantly decreased leakage rate signal.