Configuration (lds2010-compatible) – INFICON LDS3000 Modular Leak Detector User Manual
Page 24
24
Operation
6.2.4
Configuration (LDS2010-compatible)
LDS2010
setting in
menu item
22
Analog
output
LDS2010 function
LDS3000 function
Decade
scale
Upper limit
value
(10 V = ...)
1
1
Leak rate mantissa in used unit
1 ... 10 V
Leak rate mantissa
irrelevant irrelevant
1
2
Leak rate exponent (step function)
in used unit
1 ... 10 V, 0.5 V/decade,
1 V = 1E-12
Leak rate exponent
irrelevant irrelevant
2
1
Leak rate log. in used unit
1 ... 10 V, 0.5 V/decade,
1 V = 1E-12
Leak rate log.
0.5 V/dec. 1E6 [used unit]
2
2
Pressure p1 log. in used unit
1 ... 10 V, 0.5 V/decade,
1 V = 1E-3 mbar
Pressure p1
irrelevant irrelevant
3
1
Leak rate mantissa in mbar·l/s
1 ... 10 V
Leak rate mantissa
irrelevant irrelevant
3
2
Leak rate exponent (step function)
in mbar·l/s
1 ... 10 V, -1 V/decade,
0 V = 1E0 mbar·l/s
LR exponent inverted
irrelevant irrelevant
4
1
Leak rate log.
0 ... 10 V, 1 V/decade,
0 V = 1E-10 mbar·l/s
Leak rate log.
1 V/dec.
1.00E+00
4
2
Pressure p1 log. in mbar
1 V/decade, 2.5 ... 8.5 V,
2.5 V = 1E-3 mbar,
5.5 V = 1E0 mbar
p1 1 V/dec.
irrelevant irrelevant
5
1
Leak rate mantissa in used unit
1 ... 10 V rise, 0.7 ... 10 V fall
LR mantissa hyst.
irrelevant irrelevant
5
2
Leak rate exponent in used unit
1 ... 10 V, 0.5 V/decade,
0 V = 1E-14
Leak rate exponent
irrelevant irrelevant
6
1
Leak rate log. in Pa·m³/s
0 10 V, 1 V/decade,
0 V = 1E-12 Pa·m³/s =
1E-12 mbar·l/s
Leak rate log.
1 V/dec.
1E-2 mbar·l/s
6
2
Pressure p1 log. in Pa
1 V/decade, 2.5 ... 8.5 V,
2.5 V = 1E-3 mbar
p1 1 V/dec.
irrelevant irrelevant
8
1
Leak rate log. in Pa·m³/s
0 ... 10 V, 1 V/decade,
0 V = 1E-12 Pa·m
3
/s =
1E-12 mbar·l/s
Leak rate log.
1 V/dec.
1E-2 mbar·l/s
8
2
Pressure p2 log. in Pa
1 V/decade, 2.5 ... 8.5 V,
2.5 V = 1E-3 mbar
p2 1 V/dec.
irrelevant irrelevant