Starting/stopping the measurement, Loading and storing parameters, Select operation mode – INFICON LDS3000 Modular Leak Detector User Manual
Page 30: 5 starting/stopping the measurement, 6 loading and storing parameters, 7 select operation mode
30
Operation
6.5
Starting/stopping the measurement
Starting
Device in S
TANDBY
► Start measurement.
Device switched to M
EASURE
state.
Stopping
Unit measures
► Stop measurement.
Device switches to S
TANDBY
state.
6.6
Loading and storing parameters
You can use a USB stick to backup and restore the control unit and mass
spectrometer module parameters.
► Load parameters as needed:
F
UNCTIONS
> D
ATA
> P
ARAMETER
> L
OAD
> L
OAD
PARAMETER
► Save parameters as needed:
F
UNCTIONS
> D
ATA
> P
ARAMETER
> S
AVE
> S
AVE
PARAMETER
6.7
Select operation mode
V
AC
= Vacuum mode
S
NIFF
= Sniff mode
► Select operation mode as needed:
S
ETTINGS
> S
ET
UP
> O
PERATION
MODES
> V
AC
/S
NIFF
> O
PERATION
MODE
During the measurement
During standby
Z
ERO
is possible.
Z
ERO
is not possible.
The trigger outputs switch depending on
the leakage rate and the trigger threshold.
The output at the trigger outputs is:
Leakage rate value exceeded
threshold.
Sniff is possible.
Sniff is not possible.
Calibration using PLC input or output will
start an external calibration.
Calibration using PLC input and output
will start an internal calibration.