10) Press and hold the single UMX-2 button, located on the top end cap, until
the top level menu options begin scrolling on the alpha display. Once this
occurs, the button can now be released. The menu options will scroll one to
the next in a time delayed sequence, and will display all the menu options in
a continuous loop.
11) When UMX-2 is displayed, immediately press and release the button to
enter the sub menu options. Again, the options will begin scrolling in a
continuous loop (LT ON, LT OFF, LAUTO, MEM ON/OFF, UPGRAD).
12) When UPGRAD is displayed, immediately press and release the button to
set put the UMX-2 into upgrade mode.
13) Open the temporary folder with the unzipped firmware files, and double click
the batch file uflash.bat.