8 clearing a stored measurement, Clear location – Dakota Ultrasonics UMX-2 User Manual
Page 58
UMX-2 Underwater Material & Coating Thickness Gauge
54
Note: This section assumes the UMX-2 is powered up and ready to go. Also,
the UMX-2 will always return to the left most top level menu item, or the current
log file location. This was done to make it convenient for the operator to store
measurements without having to scroll through directories.
1) Press and release the single UMX-2 button, located on the top end cap, to
store a measurement, and auto increment to the next storage location.
Example: current grid cell location is A1, and auto increment is set to south.
Once the reading is stored, the cursor will automatically advance to A2.
2) It’s as simple as that!
Important Note: If the UMX-2 is currently in a stored location, the gauge will
not measure. The measurement must be cleared first, or the user must be in
an ‘empty’ storage location.
11.8 Clearing a Stored Measurement
The following procedure outlines the steps to clear a cell or log location, where the
user has opted to make and store another measurement for that specific location:
Clear Location
Note: This section assumes the UMX-2 is powered up, and the user has
navigated to the storage location that will be cleared.
1) Press and hold the single UMX-2 button, located on the top end cap, until
the top level menu options begin scrolling on the alpha display. Once this
occurs, the button can now be released. The menu options will scroll one to