beautypg.com

3 units (in/mm), Setting units – Dakota Ultrasonics UMX-2 User Manual

Page 38

background image

UMX-2 Underwater Material & Coating Thickness Gauge

34

3) When the desired gain setting is displayed, immediately press and release

the button to set the gain in the UMX-2. For this example, MED has been

selected.

4) Press and hold the button to escape out of the sub menu options at any

time.

8.3 Units (in/mm)

The UMX-2 can be set to either English or Metric units. The procedure below
outlines how to set the units in the UMX-2:

Setting Units

1) Press and hold the single UMX-2 button, located on the top end cap, until

the top level menu options begin scrolling on the alpha display. Once this

occurs, the button can now be released. The menu options will scroll one to

the next in a time delayed sequence, and will display all the menu options in

a continuous loop.

2) When UNITS is displayed, immediately press and release the button to

enter the sub menu options. Again, the options will begin scrolling in a

continuous loop (IN, MM).