Troubleshooting
Eppendorf BioSpectrometer
®
fluorescence
English (EN)
The ID cannot be set.
• Error when entering the sample ID.
Different causes are possible. For
the precise cause please see the
information in the help box.
See information in the help box.
The dilution cannot be
set.
• Error when entering the dilution.
Different causes are possible. For
the precise cause please see the
information in the help box.
See information in the help box.
Calculation not possible
because of division by
zero. Absorbance result
or Formula "b"
parameter is zero.
• An absorbance result was divided
by a "zero" value during the
evaluation of a
Division type
method (
Dual wavelength method
group). This is not mathematically
permissible.
Check the reagents and samples
used and repeat the measurement.
Do not enter "zero" as a value for
the Formula
b parameter.
There is only one
measurement left to be
performed in this series
of measurement.
The maximum number of
measurements within
one series of
measurements has been
reached.
• The number of measurements in
one measuring series is limited to
99.
Start a new series of measurement
after maximally 99 measurements.
Invalid zoom interval!
Error in the
process results method
step in the Zoom mode.
Permissible zoom range for the
wavelength scale:
• Wavelength interval at least 10 nm
• Entries for wavelengths only within
the range programmed in the
parameters for the method.
Permissible zoom range for the
absorbance scale:
• Absorbance interval at least 0.02 A
• Upper and lower limit for
absorbance interval +3 A or –3 A
Please observe the stated limits in
the zoom procedure.
Problem
Cause
Solution