Troubleshooting

Eppendorf BioSpectrometer

®

fluorescence

English (EN)

The ID cannot be set.

• Error when entering the sample ID.

Different causes are possible. For
the precise cause please see the
information in the help box.

 See information in the help box.

The dilution cannot be
set.

• Error when entering the dilution.

Different causes are possible. For
the precise cause please see the
information in the help box.

 See information in the help box.

Calculation not possible
because of division by
zero. Absorbance result
or Formula "b"
parameter is zero.

• An absorbance result was divided

by a "zero" value during the
evaluation of a

Division type

method (

Dual wavelength method

group). This is not mathematically
permissible.

 Check the reagents and samples

used and repeat the measurement.

 Do not enter "zero" as a value for

the Formula

b parameter.

There is only one
measurement left to be
performed in this series
of measurement.
The maximum number of
measurements within
one series of
measurements has been
reached.

• The number of measurements in

one measuring series is limited to
99.

 Start a new series of measurement

after maximally 99 measurements.

Invalid zoom interval!

Error in the

process results method

step in the Zoom mode.
Permissible zoom range for the
wavelength scale:

• Wavelength interval at least 10 nm
• Entries for wavelengths only within

the range programmed in the
parameters for the method.

Permissible zoom range for the
absorbance scale:

• Absorbance interval at least 0.02 A
• Upper and lower limit for

absorbance interval +3 A or –3 A

 Please observe the stated limits in

the zoom procedure.

Problem

Cause

Solution