beautypg.com

1 cleaning the cuvette shaft cover, 2 disinfection/decontamination, Cleaning the cuvette shaft cover – Eppendorf BioSpectrometer fluorescence User Manual

Page 66: Disinfection/decontamination

background image

Maintenance
Eppendorf BioSpectrometer

®

fluorescence

English (EN)

66

8.1.1

Cleaning the cuvette shaft cover

If you would like not only to clean the directly accessible surface of the cuvette shaft cover, you can remove
the cover.

1. Lift the cuvette shaft cover with one hand.

3. Clean the cover with a cloth or lint-free cotton swab dampened with a mild cleaning agent.

4. Slide the locking pin back into the housing as far as it will go.

The locking pin has completely disappeared in the housing.

8.2

Disinfection/Decontamination

1. Clean the device with a mild cleaning agent before the disinfection (see Cleaning on p. 65).

2. Choose a disinfection method which corresponds to the legal regulations and guidelines in place for

your range of application.

3. For example use alcohol (ethanol, isopropanol) or other alcoholic disinfectants.

4. Wipe the surfaces with a cloth which you have moisturized with a disinfectant.

 Do not soak the cuvette shaft cover in cleaning agent.
 Clean the cuvette shaft cover as described.

2. With the other hand, hold the cover at the height

of the locking pin and pull the cover to the right
until the locking pin has been removed.

• Pull the cover to the right at a 90 degree angle.

When the photometer is not being used, close the cuvette shaft using the blue cuvette shaft
cover to protect it from dust and other contamination.

DANGER! Electric shock as a result of penetration of liquid.

 Switch off the device and disconnect the power plug before starting cleaning or

disinfection work.

 Do not allow any liquids to penetrate the inside of the housing.
 Do not spray clean/spray disinfect the housing.
 Only plug the device back in if it is completely dry, both inside and outside.