Troubleshooting
Eppendorf BioSpectrometer

®

fluorescence

English (EN)

74

The measuring results are
not correct.

• The method has not been

programmed correctly.

 Ensure that the method parameters are

entered correctly.

• The standard solution has

not been prepared
correctly.

 Ensure that the correct standard is used

and that the measuring solution for the
standard is prepared correctly.

• The absorbance of the

reagent is drifting.

 For instable reagent absorbance and end

point methods: When measuring a long
series of samples, measure the reagent
blank value not only at the beginning but
also during the sample series. If the
reagent blank value drifts strongly, the
reagent is not appropriate for error-free
measurements and has to be replaced by a
new reagent.

• The cuvette is not

positioned correctly.

 Position the cuvette in the cuvette shaft so

that the optical window points towards the
direction of the light path.

 Photometry light path: from back to front
 Fluorimetry light path: from right to left

Error

Possible cause

Remedy