beautypg.com

4 installation, 1 preparing installation, 2 selecting the location – Eppendorf BioSpectrometer fluorescence User Manual

Page 17: 3 connecting the device to the mains, Installation 4.1, Preparing installation, Selecting the location, Connecting the device to the mains

background image

17

Installation

Eppendorf BioSpectrometer

®

fluorescence

English (EN)

4

Installation

4.1

Preparing installation

 Keep the transport carton and the packing material for subsequent safe transport or storage.

 Check the completeness of the delivery using the information in the delivery package (see Delivery

package on p. 11).

 Check all parts for any transport damage.

4.2

Selecting the location

Select the location for the BioSpectrometer fluorescence according to the following criteria:

• 2 grounded sockets for the BioSpectrometer fluorescence and for the printer.
• Solid laboratory bench with horizontal work surface

Space requirement of the device: 50 cm (with printer: 75 cm) width, 50 cm depth.

• Temperature: 15°C to 35°C.
• Avoid temperature fluctuations (e.g, caused by open windows).
• Avoid direct sunlight.
• Humidity: 25% to 70% relative humidity.

4.3

Connecting the device to the mains

1. Place the BioSpectrometer fluorescence on a suitable work surface.

2. Verify that the mains/power supply voltage and mains/power frequency match the information on the

name plate.

3. Connect the device to the mains/power line and switch it on with the power switch.

4. Remove the protective film from the display.

Ensure that no objects (e.g., loose sheets, notebooks) that could impede the flow of air are
positioned under the device.