4 replacing fuses, 5 decontamination before shipment, Replacing fuses – Eppendorf BioSpectrometer fluorescence User Manual
Page 72: Decontamination before shipment

Maintenance
Eppendorf BioSpectrometer
®
fluorescence
English (EN)
72
• Determination of the random error of the excitation light ("noise")
• Determination of the signal level and deviation of the emitted light
Select the Perform self-test function in the Device calibration group and confirm with enter.
At the end of the self-test, the display shows the message
PASSED.
If the display shows the message
FAILED, the self-test has failed. If this error cannot be corrected (see
Error messages on p. 75), contact Eppendorf Service.
8.4
Replacing fuses
The fuse holder is located between the mains connection socket and the mains power switch.
1. Disconnect the power plug.
2. Press the upper and lower end of the plastic springs
1 together and pull the fuse holder 2 fully out.
3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail
3 is positioned
correctly.
8.5
Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer
for disposal please note the following:
DANGER! Electric shock.
Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.
WARNING! Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. You find it as a PDF file on our
website (www.eppendorf.com/decontamination).
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
1
2
3