3 safety, 1 intended use, 2 user profile – Eppendorf BioSpectrometer fluorescence User Manual
Page 13: 3 warnings for intended use, 1 personal injury, Safety 3.1, Intended use, User profile, Warnings for intended use 3.3.1, Personal injury

13
Safety
Eppendorf BioSpectrometer
®
fluorescence
English (EN)
3
Safety
3.1
Intended use
The BioSpectrometer fluorescence is to be used in molecular biology, biochemistry and cell biology
research laboratories. The BioSpectrometer fluorescence is exclusively intended for use indoors. All
country-specific safety requirements for operating electrical equipment in the laboratory must be observed.
The BioSpectrometer fluorescence is used for photometric concentration determination of analytes in
liquids and recording of absorbance wavelength spectra in cuvettes. In addition, fluorescence
measurements can be used to quantify biomolecules.
Only use Eppendorf accessories or accessories recommended by Eppendorf.
3.2
User profile
The device and accessories may only be operated by trained and skilled personnel.
Before using the device, read the operating manual carefully and familiarize yourself with the device's
mode of operation.
3.3
Warnings for intended use
3.3.1
Personal injury
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device for processing any substances which could generate an explosive
atmosphere.