Methods
Eppendorf BioSpectrometer

®

fluorescence

English (EN)

36

Calibration curve

• Measurement of the RFU values and evaluation with a series of 2 to 12 standards
• Various evaluation procedures, including linear regression ("Curve fit") and nonlinear regression, can

be selected.

• Graphical and tabular display of the standard results.
• The last saved standard evaluation can be used.
• A method for calibration curve evaluation is preprogrammed.

6.4

Method parameters

This chapter illustrates the parameters for programming the methods. The order of the parameters in the
device display may slightly differ from the order in the table in order to display the parameters more clearly.
The table displays all parameters available for the various methods. Only a small portion of these
parameters are required for the corresponding method and will be shown in the display.

Parameter

Entry

Explanation

Cuvette

Selection:
10 | 5 | 2 | 1 | 0.5 | 0.2 |
0.1 mm

Optical path length of the cuvette. The device always
automatically converts absorbance values to the 10 mm path
length of a standard cuvette (see Absorbance values on p. 87).
Therefore, there is no need to change factors such as "50" for
the calculation of dsDNA concentrations when modifying the
Cuvette parameter.

No. of
wavelengths

Value input:
Range: 2 to 6.

Only for the

Multi λ method group.

Number of wavelengths at which the measurement is to be
performed.

Wavelength

Value input:
Measurement
wavelength in nm.
Range: 200 to 830 nm.

Measurement wavelength: The concentration is calculated
based on the absorbance measured with this wavelength.
For the

Multi λ and Dual wavelength method groups, enter

more than one wavelength. For some method groups (e.g.,
Nucleic acids and Proteins direct UV), the wavelengths are
preprogrammed.
For the

Dye labels method group the measuring wavelengths

are not entered into the method procedure one by one. They
are automatically imported simply by selecting the biomolecule
and dye from the

General Method Parameters function.

Wavelength
(em)

Selection:
520 nm | 560 nm

Only for the

Basic fluorimetric method group:

Measurement wavelength: The concentration is calculated
based on the fluorescence measured with this wavelength
(emission wavelength).
For the methods of the

Routine group the wavelength is

preprogrammed.

Wavelength (ex) No entry possible.

Wavelength: 470 nm

Only for fluorimetric method groups:
Excitation wavelength 470 nm is displayed.