Median filter, Dosing filter setting, Edian filter – RADWAG PUE C41H Indicator User Manual
Page 26: Osing filter setting
13.2. Median filter
This filter eliminates short mechanical shocks.
Procedure:
• Enter
NO - filter disabled
YES - filter enabled
Return to weighing:
See 11.3.
13.3. Dosing filter setting
In PUE 41 terminals an special averaging filter for dosing process has
been implemented. The result of this filtration, instead of traditional filters
for static weighing, is compared with dosing setpoints. The filter parameter
is the number of samples from the A/D converter (1 to 10). When the filter
is set to 1 every reading from the A/D converter is compared with the
dosing setpoints, which does not introduce any delay. If the filter
parameter is set to n>1, the filtering result will be calculated as
an arithmetic average from the last n measurements.
∑
=
=
n
1
i
i
X
M
,where M is a filtering result from samples X
1
to X
n .
- 26 -