Technical parameters, Echnical parameters – RADWAG PUE C41H Indicator User Manual
Page 151
31. TECHNICAL PARAMETERS
Technical data:
PUE C41H
Display LCD
Casing stainless
steel
Keyboard membrane
Power supply
85-265V AC 50/60Hz,
battery 6V 3Ah
Average operation when supplied from batteries
to 9h (buffer power supply)
Total memory size for data base
4 MB
Maximal quantity of divisions from converter
8 388 608
OIML class
III
Quantity of verifying units
6 000
Maximal increase of signal
19mV
Maximal voltage on 1 verifying unit
3,3 µV
Minimal voltage on 1 verifying unit
1µV
Minimal impedance of load cell
90
Maximal impedance of load cell
1200
Power voltage of load cell
5V
Connection of load cells
4 or 6 cables + screen
Working temperature
-10°C to +40°C
IP rating
IP 66/67
Optoinsulated interfaces
RS 232 i RS 485
Inputs / outputs
3 optoinsulated inputs
3 optoinsulated reed
relay outputs
- 151 -