Drawing of cables and outputs, We 4 - 4 inputs / 4 outputs module, Technical specification – RADWAG PUE C41H Indicator User Manual
Page 125: Echnical specification
27.3.3. Drawing of cables and outputs
Relay outputs diagram:
SIGNALS AND DESIGNATIONS
OF CONDUCTORS
Wire number
Description
1 OUT
1
2 OUT
2
3 OUT
3
4 OUT
4
5
(yellow -
green)
Common
27.4. WE 4 - 4 inputs / 4 outputs module
WE 4 module comprises 4 optoinsulated inputs and 4 optoinsulated
outputs of reed relays. The input / output wires are led out via a gland
on the back wall of the housing (3m length).
Caution:
As standard indicator is equipped with 3 in and 3 out sockets.
27.4.1. Technical specification
Parameters of outputs
Quantity of outputs
4
Type of outputs
Reed operation contacts
Wire diameter
0,14 - 0,5mm
2
Maximal load-current contact
capacity
0,2A DC
Maximal forward voltage
50V DC
Parameters of inputs
Quantity of inputs
4
Input type
Optoinsulated
Wire diameter
0,14 – 0,5mm
2
Control voltage range
5 -24V DC
- 125 -