Error messages, Trouble shooting, Rror messages – RADWAG PUE C41H Indicator User Manual
Page 150: Rouble shooting

- 150 -
29. ERROR MESSAGES
ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING (where: XXX –
parameter name) – confirmed with ENTER of a wrong value in the user’s
menu,
Err2
- Value beyond the zero range,
Err3
- Value beyond the tare range,
Err4
- Calibration mass or start mass adjustment error
(
±1% for weight, ±10% for start mass),
Err5
- Mass of a single piece lower than the 0.1 of scale division,
Err6
- Mass of all pcs (declared as standard) lower than one
division,
Err8
- Tarring / zeroing operation time exceeded,
NULL
- Zero value from the AD converter,
FULL2
- Measurement
range overflow,
HI
- Display range overflow,
LH
- Start mass error, the mass on the weighing platform
is beyond the acceptable range
±20 of start mass
30. TROUBLE SHOOTING
Problem Cause
Solution
Scale switches
off automatically
„t1” parameter set to
„YES” (Power save)
In
„LH” appears on
the display
Weight pan not empty
when switching on
Unload the pan. Indication returns
to zero