29. ERROR MESSAGES


ERROR ”XXX” ESC – RETURN TO PREVIOUS SETTING (where: XXX –
parameter name) – confirmed with ENTER of a wrong value in the user’s
menu,

Err2

- Value beyond the zero range,

Err3

- Value beyond the tare range,

Err4

- Calibration mass or start mass adjustment error

(

±1% for weight, ±10% for start mass),

Err5

- Mass of a single piece lower than the 0.1 of scale division,

Err6

- Mass of all pcs (declared as standard) lower than one

division,

Err8

- Tarring / zeroing operation time exceeded,

NULL

- Zero value from the AD converter,

FULL2

- Measurement

range overflow,

HI

- Display range overflow,

LH

- Start mass error, the mass on the weighing platform

is beyond the acceptable range

±20 of start mass



30. TROUBLE SHOOTING

Problem Cause

Solution

Scale switches
off automatically

„t1” parameter set to
„YES” (Power save)

In change
to „NO”

„LH” appears on
the display

Weight pan not empty
when switching on

Unload the pan. Indication returns
to zero