28.4.12. Switch on continuous transmission in current unit

Syntax: CU1 CR LF

Possible answers:


CU1_I CR LF

- command comprehended but cannot be executed

CU1_A CR LF
MASS FRAME

- command comprehended and in progress

- mass value in current unit is returned


Frame format:

1 2 3

4

5

6 7-15 16 17 18 19 20 21

S U I stability space sign mass space

unit

CR LF



28.4.13. Switch off continuous transmission in current unit

Syntax: CU0 CR LF

Possible answers:

CU0_I CR LF

- command comprehended but cannot be executed

CU0_A CR LF

- command comprehended and executed



28.4.14. Lock the scale keyboard

Syntax: K1 CR LF

Possible answers:

K1_I CR LF

- command comprehended but cannot be executed

K1_OK CR LF

- command executed


Caution: