Continuous transmission, Configuring printouts, Ontinuous transmission – RADWAG PUE C41H Indicator User Manual
Page 149

- 149 -
Example 2:
? _ - _ _ _ _ 2 . 2 3 7 _ l b _ CR LF - the printout generated from 
the scale after pressing ENTER/PRINT. 
 
 
Example 3:
^ _ _ _ _ _ _ 0 . 0 0 0 _ k g _ CR LF - the printout generated from 
the scale after pressing ENTER/PRINT. 
 
 
 
28.6. Continuous transmission
The indicator can work in a continuous transmission mode. It can be 
switched on or off in parameters or using RS232 commands. 
Frame format sent by the indicator when continuous transmission is set:
1
2
3 4
-12 13 14 15 16 17 18
stability space
sign
mass space
unit
CR LF
Stability character
[space] if stable 
[?] if not stable 
[^] if an indication over the range 
[v] if fan indication below the range
sign
[space] for positive values or 
[-] for negative values
mass
9 characters justified to the right
unit
3 characters justified to the left
command
3 characters justified to the left
 
 
 
28.7. Configuring printouts 
 
If some information included are redundant or not sufficient and there is 
a necessity of changes you can design a non-standard printout. There is 
possible to create up to four standard printouts (see the chapter. 17). 