Set tare value, Send the stable result in basic unit, Et tare value – RADWAG PUE C41H Indicator User Manual
Page 141: End the stable result in basic unit

- 141 -
28.4.4. Set tare value 
 
Syntax: UT_TARE CR LF, where TARE – tare value 
 
Possible answers: 
 
UT_OK CR LF
- command executed
UT_I CR LF
- command comprehended but cannot be executed
ES CR LF
- command not recognised (possible wrong tare format)
 
Notice: 
This protocole uses the dot character as a decimal point for separating 
the decimal fraction part. 
 
 
 
28.4.5. Send the stable result in basic unit 
 
Syntax: S CR LF 
 
Possible answers: 
 
S_A CR LF 
S_E CR LF 
- command accepted and in progress 
- time limit for stable result exceeded 
S_I CR LF
- command comprehended but cannot be executed
S_A CR LF 
MASS FRAME 
- command accepted and in progress 
- mass value in basic unit is returned 
 
Frame format: 
1 2-3
4
5 6 7-15 16 17
18
19
20
21
S space stability space sign
mass
space unit CR
LF
 
 
Example:
S CR LF – computer command 
S _ A CR LF - command accepted and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF – command done, mass value 