beautypg.com

Switch on continuous transmission in basic unit, Switch off continuous transmission in basic unit, Witch on continuous transmission in basic unit – RADWAG PUE C41H Indicator User Manual

Page 144: Witch off continuous transmission in basic unit

background image

- 144 -

Possible answers:

SUI_I CR LF

- command comprehended but cannot be executed

SUI_A CR LF
MASS FRAME

- command accepted and in progress
- mass value in current unit is returned immediately


Frame format:

1 2 3

4

5

6 7-15 16 17 18 19 20 21

S U I stability space sign mass space

unit

CR LF


Example:

S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned



28.4.10. Switch on continuous transmission in basic unit

Syntax: C1 CR LF

Possible answers:

C1_I CR LF

- command comprehended but cannot be executed

C1_A CR LF
MASS FRAME

- command comprehended and in progress

- mass value in basic unit is returned


Frame format:

1 2

3

4

5

6 7-15

16 17 18 19 20 21

S I space stability space sign mass space

unit

CR LF



28.4.11. Switch off continuous transmission in basic unit

Syntax: C0 CR LF

Possible answers:

C0_I CR LF

- command comprehended but cannot be executed

C0_A CR LF

- command comprehended and executed