HOLD B.
NO
CHUTE ENABLE
NO
• Enter the submenu
according the table below:
MIN NO
OK NO
MAX NO
STABLE NO
THRESHOLD 1
OUT1
THRESHOLD 2
OUT2
CHUTE OUT3
2. Return to weighing according to ch.11.3.
3. Connect the cable Scale - I/O (see ch. 25) and then install external
buttons START, STOP (terminal inputs) and subsequent devices
to the terminal outputs according to the drawings below:
4. Enter
the
DOSING operating mode according to ch. 24.8.6,
5. Press
to enter thresholds setting:
• MIN – mass value over which the bulk dosing changes to the precise
dosing ,
• MAX – mass value over which dosing is completed and CHUTE starts.
6. Before you start dosing unload the pen (container, silo) or tare the
scale pressing
.
7. Start the process by pressing the external START button,
pictogram starts to blink, and threshold 1, THR 1 bulk dosing,
is performed.