Read lower threshold, Read upper threshold, Send all implemented commands – RADWAG PUE C41H Indicator User Manual
Page 147: Ead lower threshold, Ead upper threshold, Possible answers, Syntax: pc cr lf

- 147 -
Possible answers:
UH_OK CR LF
- command executed
ES CR LF
- command not comprehended (wrong mass format)
28.4.20. Read lower threshold
Syntax: ODH CR LF
Possible answers: DH_MASA CR LF - command executed
Frame format:
1 2
3
4-12
13 14 15 16
17
18 19
D H space mass space unit space
CR
LF
Mass - 9 characters justified to the right
Unit
- 3 characters justified to the left
28.4.21. Read upper threshold
Syntax: OUH CR LF
Possible answers: UH_MASA CR LF - command executed
Frame format:
1 2
3
4-12
13 14 15 16
17
18 19
U H space mass space unit space
CR
LF
Mass - 9 characters justified to the right
Unit
- 3 characters justified to the left
28.4.22. Send all implemented commands
Syntax: PC CR LF