- 69 -

Revision 4.6 3/2004

Step 31

With the applied pressure displayed press the

enter key.

ENT

The calibrator will respond by displaying:

+xxxxx working

After 3-5 seconds the calibrator will request the next
calibration point by displaying.

Apply and enter

value 2.00000

Step 32

Apply the pressure required for the next cali-

bration point and enter the applied pressure value with
the number keys. For the second point a pressure of
20.003 has been applied.

Apply and enter

value 20.003

Step 33

With the required pressure level applied and

shown on the display press the enter key.

ENT

Step 34

Repeat steps 23-26 as many times as required

to produce the desired number of linearization points.
Points must be in sequence (low to high).

A maximum of 18 linearization points plus the zero and
span points may be programmed for Quick Select pres-
sure modules. Factory calibration consists of a 10 point
linearization combined with a 10 point upscale and 10
point down scale linearization and hysteresis check.

Important Note:

The number of calibration points

needed to assure adequate correction for non-linearity is
as follows:

Sensor Accuracy number of Points

0.025% & 0.05%

20

including temperature

(5% increments)

0.025-0.07%

11

(10% increments and zero)

0.1%

6

(20% increments and zero)

0.25%

5

(25% increments and zero)

Step 35

When the final linearization point to be per-

formed has been completed press the clear entry key.

CE

The calibrator will respond by displaying:

Calibration done

Save

Cancel

Step 36

To save the new linearization data to memory,

thereby overwriting the previously stored values in the
EEPROM, select “Save” and press enter. To discard the
newly generated calibration information select “Cancel”
and press enter.

Step 37

If “cancel” is selected the calibrator will exit

the calibration mode. If the enter key is pressed with the
word “save” flashing the calibrator will prompt entry of
the date of the new calibration as follows:

mmddyy

- - - - - -

Step 38

Using the number keys enter the current date

such as:

mmddyy

031196

For March (03), eleventh (11), 1996

Step 39

With the desired date entered and displayed,

press the enter key.

ENT

Step 40

Turn off the calibrator and install the desired

Quick Select module.

Step 41

Turn on the calibrator to resume general pres-

sure measurement activities.