Linearity Cal

Yes No

Step 18

If calibration activity is to be limited to zero and

span, using the arrow keys select the word no (when
selected the word “no” will flash on the display). If lin-
earity adjustment is to be performed proceed to Step 21.

Linearity Cal

Yes No

Step 19

With the word no flashing press the enter key.

ENT

The calibrator will respond by displaying:

Calibration done

Save Cancel

Step 20

To save the new zero and span values to

memory, thereby overwriting the previously stored
values in the EEPROM, select the word “Save” and
press the enter key. To discard information and retain the
information currently in the EEPROM of the calibra-
tor select “Cancel” and press the enter key . If the save
option is selected proceed to Step 21. To re-linearize
module proceed to Step 26.

Step 21

If “cancel” is selected the calibrator will exit

the calibration mode. If “save” is selected the calibrator
will prompt you to enter the date of the new calibration
as follows:

mmddyy

- - - - - -

Step 22

Using the number keys enter the current date

such as:

mmddyy

031196

For March (03), eleventh (11), 1996

Step 23

With the desired date entered and displayed,

press the enter key.

ENT

Step 24

Turn off the calibrator and install the desired

Quick Select module.

Step 25

Turn on the calibrator to resume general pres-

sure measurement activities.

Step 26

To re-linearize a Quick Select pressure module

select the word “Yes” to the linearity query.

Important Note:

If full recalibration of the pressure

module is desired the zero and span adjustment must
be reset through these procedures prior to entering the

linearization mode. The zero and span values established
in these procedures will be used for the zero and span
values of the new calibration data. If zero and span are
not reset at this time the recalibration will use the values
for zero and span stored in memory from the previous
zero and span activities.

Linearity Cal

Yes No

Step 27

With the word “Yes” flashing press the enter key.

ENT

The calibrator will respond by displaying:

Clr Old Lin Data

Cancel Proceed

Step 28

Use the arrow keys to select “Proceed” to con-

tinue the linearization process. Select “Cancel” using the
arrow keys to discontinue the re-linearization process.

***WARnInG***

Entering “Proceed” will erase linearization data in the
EEPROM of the calibrator.

Clr Old Lin Data

Cancel Proceed

Step 29

To continue the linearization process, with the

word “Proceed” flashing press the enter key.

ENT

The calibrator will respond by displaying:

Apply and enter

value 1.00000

Step 30

Apply the pressure required for point one and

enter the numerical value, using the number keys of the
exact pressure level applied. (Use the engineering unit in
use for the left pressure module at the time the calibra-
tion process was started). When point 1 has been applied
and entered the display will look as follows for calibra-
tion point of 10.002.

Apply and enter

value 10.002