beautypg.com

Batfunc percent usereng h2oref, Auto shut off cfgbattindicator, Auto shut off disable enable – Ashcroft ATE-100 Handheld LCD Digital Calibrator User Manual

Page 11: Battery indicator disable enable

background image

- 11 -

Revision 4.6 3/2004

BatFunc PerCent

UserEng H2Oref

Step 3

With “BatFunc” flashing press the enter key.

ENT

Step 4

Using the up/down arrow keys to select Auto

Shut off. When selected the text will flash.

Auto shut off

CfgBattIndicator

Step 5

With the words “Auto shut off” flashing press

enter to activate the battery save function.

ENT

The PTE will respond by displaying the following:

Auto Shut off

Disable Enable

If Enable is selected the HHC will automatically turn
off if there is no keypad activity detected for a period of
10 minutes. If Disable is selected the HHC will remain
on continuously until the battery voltage is no longer
sufficient to power the system. The battery save mode
selected is stored in Electrically Erasable Programmable
Read Only Memory (EEPROM). As such it does not
require reprogramming on power up.

note: An optional AC adapter is available for long term
continuous measurement activities.

Section 3.4.2 Set Up for Battery Charge Level

Indication

If this function is not set up pressing the BAT CK key
will provide a measurement of the remaining voltage
of the installed batteries.

If the battery check function is set up, the HHC can
provide an estimate of the remaining battery life. This
estimate is expressed as a percentage of the expected
30 hour life of newly installed alkaline batteries. Use
of non-alkaline, lithium or rechargeable batteries will
render the life expectancy estimate inaccurate.

This

function should be set up only when new batteries
are installed.

If this function is not set up pressing the BATCK key
will provide the measured remaining voltage of the
installed batteries.

To have the capability to check the estimated remaining
battery life follow the steps below.

Step 1

Press the On/OFF key on the keypad to turn

the power on to the base unit.

ON

OFF

Step 2

With the HHC on and reading pressure press the

battery check key, designated BAT CK.

SET

UP

Step 3

Using the arrow keys select the battery save

function (designated BatFunc in the set up menu). When
selected, the text “BatFunc” will flash on the display.

BatFunc PerCent

UserEng H2Oref

Step 4

With the “BatFunc” text flashing press the enter

key.

ENT

Step 5

Using the up/down arrow keys to select CfgBat-

tIndicator to set battery level. When selected the text will
flash.

Auto shut off

CfgBattIndicator

In response the HHC will display:

Battery Indicator

Disable Enable

Step 6

To Enable the battery level indictor use the

arrow keys to select the text Enable

Step 7

With the text Enable flashing, press the enter

key.

ENT

The HHC will respond by displaying:

Set BattCapacity

XXXXXX

where:

XXXXXXXXXXXXXXXXX is a bar graph represen-
tation of the remaining battery life and +X.XXV repre-
sents the voltage level of the installed batteries.

Step 8

After installation of new batteries use the right

pointing arrow key to index the bar graph to 100%.

>

Step 9

After the bar graph has been refreshed press the

enter key.

ENT

Note:

The bar graph and the estimate of the percent bat-

tery life remaining will now be displayed along with the