Electrical pressure, Zero cal yes no, Full scale cal yes no – Ashcroft ATE-100 Handheld LCD Digital Calibrator User Manual

Page 67

background image

- 67 -

Revision 4.6 3/2004

Step 8

Enter the 5 digit calibration access code pro-

vided with the calibrator base unit.

Access code ?

12345

Step 9

With the correct access code displayed press

the enter key.

ENT

The calibrator will respond by displaying:

Electrical

Pressure

Step 10

Using the down arrow key select the word

“Pressure”. When selected, the word Pressure will flash
on the display:

Electrical

Pressure

Step 11

With the word “Pressure” flashing press the

enter key.

ENT

The calibrator will respond by displaying:

Zero Cal

Yes No

Step 12

To perform zero calibration (adjustment) select

Yes and press the enter key.

ENT

To skip zero adjustment and proceed to full scale cali-
bration (adjustment) select no and press the
enter key.

Step 13

With the pressure module at zero pressure,

press the enter key.

ENT

note: The following should be done to zero each type of
pressure module:

Pressure
Type

Zero Source

Absolute – Apply absolute zero using vacuum pump.
Gauge –

Vent to atmosphere

Compound – Vent to atmosphere
Differential – Vent both ports to atmosphere
Vacuum –

Vent to atmosphere

The calibrator will respond by displaying:

xxxxx working

Where xxxxxx is the voltage being read from the Quick

Select pressure module and the flashing word “working”
indicates that the calibrator is making the internal adjust-
ments necessary for the recalibration.

After approximately 3 seconds the calibrator will dis-
play:

Full Scale Cal

Yes No

Step 14

To perform full scale calibration (adjustment)

select Yes and press the enter key.

ENT

To skip full scale adjustment and proceed to re-lineariza-
tion process select no and press the enter key.

Important Note:

If full recalibration of the pressure

module is desired the zero and span adjustment must
be reset through these procedures prior to entering the
linearization mode. The zero and span values established
in these procedures will be used for the zero and span
values of the new calibration data. If zero and span are
not reset at this time the recalibration will use the values
for zero and span stored in memory from the previous
zero and span activities.

Step 15

Apply the full scale pressure for the Quick

Select pressure module being calibrated.

(Coming as close as possible to the full scale measure-
ment range of the pressure module being calibrated)
with the pressure applied press the enter key.

ENT

The calibrator will respond by displaying:

Enter pressure

applied .00000

Step 16

Using the keypad enter the exact value of the

pressure applied to the Quick Select pressure module.
Be certain to use the engineering unit in use for the left
Quick Select module when calibration activities com-
menced.

Enter pressure applied

100.01

In the above example a pressure of 100.01 is being
applied to the left Quick Select pressure module.

Step 17

With the full scale (F.S.) pressure applied and

the corresponding value shown on the display, press the
enter key.

ENT

The calibrator will respond by displaying: