- 34-

Revision 4.6 3/2004

ON

OFF

Step 4

To access the temperature setup menu press the

Set up key.

SET UP

Step 5

Use the downward pointing arrow to access the

following selection screen

Calib RTmodule

dp Meas* TCmodule

* dp Meas is an optional feature that allows for the
subtraction or addition of the measurement value of one
sensor from the measurement value of the other sensor.
If the calibrator was not ordered with this option it will
nOT appear on the screen.

Step 6

As required, using the left/right and up/down

arrows select the text “RTmodule”. The text RTmodule
will flash when selected.

Calib RTmodule

dp Meas

Step 7

With the text “RTmodule” flashing press the

enter key.

ENT

The calibrator will respond by displaying:

Probe Connection

2W 3W 4W

This represents the configuration of the RTD to be used,
where the number equals the number of wires the RTD
uses. note: The use of a 4 wire RTD is recommended for
optimum accuracy.

Step 8

Using the left/right arrow keys select the appro-

priate configuration for the RTD probe to be used. In the
example below the 4 wire configuration was selected.

Probe Connection

2W 3W 4W

Step 9

With the desired wiring configuration selected

press the enter key.

ENT

The calibrator will respond by displaying the available
units of measure as follows:

Engnring units:

o

C

o

F K

o

R

Step 10

Using the left/right arrow keys select the

desired engineering unit for the measurements to be
made. In the example here the measurements will be

made in degrees Celsius.

Engnring units:

o

C

o

F K

o

R

Step 11

With the desired engineering unit selected and

flashing press the enter key.

ENT

The calibrator will respond with the following display:

Probe: 12345678

P&a = 385 RO = 1000

On initial use the first selection and the corresponding
probe will appear selected. Use the arrow keys to select
the desired coefficients from the available factory pro-
grammed curves. The GQS-RT1 is pre-preprogrammed
for the following RTDs:

Position Probe

1

α38

5

–200/850

2

α392

∠200/850

3

ni Minco

(ni 120

Ω)

4

Cu 10 ohm

5-8

available for programming using the

optional calibration and programming

software for the GQS-RT1 module.

Important Note:

The GQS-RT1 is compatible with RTD

probes with output resistance values of between 0 and
400 ohms. Use of the module with higher resistance
RTD probes will result in diminished accuracy and pos-
sible damage to the module electronics.

The GQS-RT2 is pre-preprogrammed for the following
RTDs:

Position

Probe

1

α38

5

–200/850

1

PT

1000

2-8

available for programming using the optional

calibration and programming software for

the GQS-RT2 module.

Important Note:

The GQS-RT2 is compatible with RTD

probes with output resistance values of between 1000
and 4000 ohms. Whereas this module will function with
lower resistance RTDs accuracy will be greatly dimin-
ished. In addition, use of the module with higher resis-
tance RTD probes will result in diminished accuracy and
possible damage to the module electronics.

Step 12

Using the left/right arrow keys select the RTD

probe type for the measurements to be made. In the
example here the measurements will be made in using
the PT 100 type 385