beautypg.com

Evnttimr dampen leakrate status, Leak mode rate decay – Ashcroft ATE-100 Handheld LCD Digital Calibrator User Manual

Page 32

background image

- 32-

Revision 4.6 3/2004

Step 9

Using the number keys enter the time inter-

val, in seconds, over which the leak rate test is to be
performed. The time interval may be from 1 to 65,000
seconds.

Time to Monitor

sec. 12345

Step 10

With the desired test interval displayed press

the enter key.

ENT

At this point the HHC will return to basic pressure
measurement.

See Sections 15.3 and 15.4 for information on perform-
ing leak rate testing.

Section 15.2 Pressure Decay Test Function Setup

Step 1

With the HHC on and displaying pressure press

the set up key.

SET

UP

Step 2

Using the arrow keys, locate and select the

words “LeakRate” in the set up menu. When selected the
words “LeakRate” will flash.

EvntTimr Dampen

LeakRate Status

Step 3

With the words LeakRate flashing press the

enter key.

ENT

The HHC will respond by displaying:

Leak Mode

Rate Decay

Step 4

Using the left/right arrow select the word Decay.

Leak Mode

Rate Decay

Step 5

With the word Decay flashing press the enter key.

ENT

Step 6

The HHC will respond by displaying:

Time to Monitor

sec. xxxxx

Step 7

Using the number keys enter the time interval,

in seconds, over which the pressure decay test is to be
performed. The time interval may be from 1 to 65,000
seconds.

Time to Monitor

sec. 12345

Step 8

With the desired test interval displayed press

the enter key.

ENT

At this point the HHC will return to basic pressure
measurement.

See Sections 15.3 and 15.4 for information on perform-
ing leak rate testing.

Section 15.3 Sensor Selection; Rate or Pressure

Decay Tests

In order to perform leak testing the desired Quick Select
pressure module must be monitored by the HHC system.
The following provides information on activating the
desired pressure module for leak test applications:

With Two Quick Select Pressure Modules Installed

If two pressure modules are installed and the output of
both are displayed, the HHC will default to the left pres-
sure module.

If two Quick Select pressure modules are installed and
one of the two modules is deselected, meaning the
measurement values of that module are not shown on the
display, leak testing will use the active pressure module
to perform the leak test measurements.

The port select key (PORT SEL) is used to select or
deselect a module. When a pressure module is dese-
lected the corresponding side of the display will indicate
one of the following: mA, Volts or “------“.

If only one Quick Select pressure module is installed the
HHC will default to the installed pressure module for
leak test measurements.

Section 15.4

Performing Leak Rate or Pressure

Decay Tests

Leak rate and pressure decay tests can be performed as
follows:

Step 1

With the HHC on and reading pressure connect

the pressure port of the selected pressure module to the
vessel to be monitored. If the pressure module incorpo-
rates a differential pressure sensor use the High pressure
port for the test procedure.

Step 2

For Pressure Decay Select the desired engi-

neering unit to be used as described in Section 8.

For Leak Rate All leak rate measurements are made in
cc/seconds and require no selection at this point.

Step 3

Pressurize the vessel to the required level for

the test and allow vessel pressure to stabilize from the
effects of sudden pressure increase or decrease.

Step 4

Press the key designated LEAK on the keypad