Appendix 5: Digital current loop module

M210282en-A

90

2.

OPERATION

The HMP228 transmitters can either be given an address or operated without
an address. Both single and dual loop wiring with half duplex connection can
be used. No address is needed when only one HMP228 transmitter is used;
when several transmitters are connected to the same line, each transmitter
must be given an address in the initial configuration.

A current loop must get its operating voltage from the master or it can have its
own (floating) power supply capable of supplying 15...40 V and 20...30 mA.
Unregulated AC/DC adapter can be used, if the current is limited to 20 mA at
least by a serial resistor.

Note.The host computer can restrict the loop supply voltage that can be used;
see computer specifications.

The serial line structure is a serial interfaced chain (daisy chain). At one end
of the serial line there must be a HMP228 transmitter and at the other end a
line master. A branch line can be made with a junction box.

TX HI

TX LO

TWISTED PAIR WIRING

-

HOST COMPUTER

DUAL LOOP WIRING

TWISTED PAIR WIRING

HOST COMPUTER

RX LO

RX HI

TX HI

TX LO

RX LO

RX HI

TX HI

TX LO

RX LO

RX HI

TX HI

TX LO

RX LO

RX HI

TX HI

TX LO

RX LO

RX HI

120R

33 nF

TX HI

TX LO

RX LO

RX HI

120R

33 nF

120R

33 nF

NOTE: DYNAMIC LINE TERMINATION
REQUIRED IF NO DEVICE
AT THE END OF THE BUS.

NOTE: DYNAMIC LINE TERMINATION
REQUIRED IF NO DEVICE
AT THE END OF THE BUS.

HMP228
transmitters

1 to n pieces

HMP228
transmitters

1 to n pieces

SINGLE LOOP WIRING