Checking the serial port network operation – Vaisala HMP228 User Manual

Page 91

background image

HMP228

M210282en-A

Appendix 4: RS 485/422 serial port module

85

4.

CHECKING THE SERIAL PORT NETWORK OPERATION

Normally measurement readings are asked when the transmitter is in POLL
mode; then the command must include the address of the transmitter. If the
settings of the transmitter need to be changed, the transmitter is switched to
STOP mode with command OPEN; commands can then be given without
address. When the line to the transmitter is closed, it returns to POLL mode.

STOP mode

Open the line to the transmitter:

open 22
HMP 22 line opened for operator commands

Transmitter no. 22 is now temporarily set to STOP mode; it accepts com-
mands sent without address until CLOSE command is given. Individual set-
tings can now be easily modified. Do not open more than one line at a time.

Use command ? to find out the settings of the active transmitter:

?
HMP228 / 1.01
CPU serial nr : 0
Keyboard type : 0
Address

: 7

Output units

: metric

Baud P D S

: 4800 E 7 1 FDX

Serial mode

: STOP

Output intrv. : 0 min
Mtim

: 32

Pressure : 1013.25
Analog outputs

Ch1 0.00 ... 10.00 V

Ch2 0.00 ... 10.00 V

Ch1 ( aw )

lo

0.000

Ch1 ( aw )

hi

1.000

Ch2 ( T )

lo

-20.000 'C

Ch2 ( T )

hi

180.000 'C

Transducer

:

PRB serial nr : 0
Calibr. date

: 0

When the necessary settings have been given, close the line to transmitters
(the command closes all open lines):

>close
line closed

CLOSE command is always given without address. If no lines are open, there
will be no response to the CLOSE command.