HMP228 Transmitters

M210282en-A

Appendix 1: Serial commands

47

For example, water activity is selected to be output on channel 1 and tem-
perature on channel 2; the temperature range is scaled to 0...100

°

C:

>ASEL RH T
Ch1

(aw) lo 0.000

?

Ch1

(aw) hi 1.000

?

Ch2

(T ) lo -40.000 'C

? 0

Ch2

(T ) hi +160.000 'C

? 100

When the ASEL command is given on its own, the transmitter outputs its cur-
rent settings:

>ASEL
Ch1

(aw) lo

0.000

Ch1

(aw) hi

1.000

Ch2

(T ) lo

-40.000 'C

Ch2

(T ) hi

+160.000 'C

The outputs and their scales can also be given directly with the ASEL com-
mand.

ASEL xxx yyy aaa.a bbb.b ccc.c ddd.d

xxx

=

channel 1's quantity

yyy

=

channel 2's quantity

aaa.a

=

lower limit of channel 1

bbb.b

=

upper limit of channel 1

ccc.c

=

lower limit of channel 2

ddd.d

=

upper limit of channel 2

ASCL

Scaling the analogue outputs

Disconnect the security lock jumper!

ASCL

Scales the outputs selected on channels 1 and 2.

For example, scaling water activity on the range of 0...1.00 and temperature -
40...+160 °C:

>ASCL
Ch1

(aw) lo

0.000

?

Ch1

(aw) hi

1.000

?

Ch2

(T ) lo

0.000 'C

? -40

Ch2

(T ) hi

100.000 'C

? 160