Check that the reading corresponds with the reading of the reference
instrument. If not, adjust the second point.

If necessary, adjust with the arrow switches to the correct value and
press ENT. At the second calibration point the second LED from the
left flashes.

4.6.3

Temperature channel adjustment with Pt 100 simulators

Switch the power off and disconnect the wires to the Pt 100 sensor from solder
lugs TP5, TP6 and TP7.

TP6

TP5

TP7

X88

Figure 4.6.3.1

Location of solder lugs TP5, TP6 and TP7 and

connector X88

Connect a Pt 100 simulator to connector X88 and set it at the lowest tempera-
ture to be calibrated.

X88

Pt 100

Figure 4.6.3.2

Connecting the Pt 100 simulator to connector X88

Switch the power on. Follow the one or two point calibration instructions in
Chapters 4.6.1 - 4.6.2.

Switch the power off. Disconnect the Pt 100 simulator and reconnect the Pt
100 wires to solder lugs TP5, TP6 and TP7.

The correct connections according to the wire colours are:

TP5

TP6

TP7

TP8

blue

green

yellow

black

If there is not a Pt 100 simulator available, the adjustment can be made with
two resistors of 84

and 154

whose resistance is known precisely. Measure

the resistor with a resistance meter. Look up the corresponding temperature
value from a Pt 100 conversion table or calculate it using the following
equation: