4.4.5

Humidity calibration table

Temperature

°C

15

20

25

30

35

°F

59

68

77

86

95

LiCl

%RH

*

11.3

11.3

11.3

11.3

4...20 mA

5.81

5.81

5.81

5.81

0...20 mA

2.26

2.26

2.26

2.26

0...1 V

0.113

0.113

0.113

0.113

0...5 V

0.565

0.565

0.565

0.565

0...10 V

1.13

1.13

1.13

1.13

NaCl

%RH

75.6

75.5

75.3

75.1

74.9

4...20 mA

16.10

16.08

16.05

16.02

15.98

0...20 mA

15.12

15.10

15.06

15.02

14.98

0...1 V

0.756

0.755

0.753

0.751

0.749

0...5 V

3.780

3.775

3.765

3.755

3.745

0...10 V

7.56

7.55

7.53

7.51

7.49

Table 1

Greenspan's calibration table

*) If the LiCl solution is used or stored in temperatures below +18 °C (+64 °F), the

equilibrium humidity of the salt solution changes permanently.

4.6

Temperature calibration

The temperature channel has been calibrated at the factory and since it is very
stable, adjustment should be made only when there is strong reason to believe
that the adjustments have changed.

Temperature calibration should be performed against some accurate
temperature reference. It can be done either using the press switches inside the
housing, through the serial bus or the menus on the local display. Either a one
point offset correction or a two point calibration is possible.

Disconnect the security lock jumper!

4.6.1

One point offset correction

4.6.1.1

Using serial commands

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Check the transmitter against the reference.

Give command CT , enter the first point value and press :

>CT
T : 0.90 Ref1 ? yy.y
Press any key when ready

If you want to see how the sensor stabilizes to the reference
temperature, enter c instead of the first reference: