Vaisala HMP228 User Manual
Page 47
HMP228
M210282en-A
Operating Manual
41
5.5
Electronics
User interface
3 keys and 4 LEDs inside the
housing or local display keypad
Display
2 x 16 character alphanumeric
high-contrast, wide view angle
LCD
character height
3.85 mm (0.15")
Keyboard
1 x 4 keypad
Connections
screw terminals, 0.5 mm
2
wires
(AWG 20), stranded wires
recommended
Operating voltage
24 VDC / isolated 24 VAC
(20...28 V)
option
115/230 VAC with power supply
module
Power consumption
100 mA maximum (24 VDC)
of the alarm relays
55 mA max (24 VDC)
Recommended external load for
current outputs
<500
Ω
0...1 V voltage output
>2 k
Ω
(to ground)
0...5 and 0...10 V voltage outputs
>10 k
Ω
(to ground)
Operating temperature (electronics)
-40...+60 °C
with display cover
0...+50 °C
with power supply module
-40...+45 °C
with alarm outputs up to 8A
-40...+45 °C
with alarm outputs up to 6A
-40...+60 °C
Storage temperature
-40...+70 °C
Pressure range of the HMP228
0...40 bar
sensor head
5.6
Mechanics
Housing material
G-AlSi12 (DIN 1725)
Housing classification
IP 65 (NEMA 4)