beautypg.com

Vaisala HMP228 User Manual

Page 46

background image

HMP228
Operating Manual

M210282en-A

40

5.

TECHNICAL DATA

5.1

Water activity

Measuring range of water activity

0...1

Accuracy (including nonlinearity and repeatability)

maximum achievable accuracy when calibrated against high quality,
when calibrated against salt solutions (ASTM E104-85):

±0.02 (0...0.9)
±0.03 (0.9...1.0)

maximum achievable accuracy when calibrated against high-quality,
certified humidity standards:

±0.01 (0...0.9)
±0.02 (0.9...1.0)

Response time (90 %) at +20 °C in
still oil (stainless steel filter)

10 min

Sensor

HUMICAP

thin film polymer sensor
(part no. 19525HM)

5.2

Temperature

Measuring range

-40...+180 °C

Typical accuracy of electronics at

± 0.1 °C

+20 °C (+68 °F)

Typical temperature dependence of

± 0.005 °C/°C

electronics

Temperature sensor

Pt 100 IEC 751 1/3 class B
(part no. 10429)

5.4

Outputs

Two analogue outputs selectable

0...20 mA

4...20 mA

0...1 V

0...5 V

0...10 V

Typical accuracy of analogue output

± 0.05 % full scale

at +20 °C
Typical temperature dependence of

0.005 %/°C full scale

analogue output
Serial output

RS232C