beautypg.com

Vaisala HMP228 User Manual

Page 54

background image

HMP228 Transmitters
Appendix 1: Serial commands

M210282en-A

48

The output scales can also be given directly with the ASCL command.

ASCL aaa.a bbb.b ccc.c ddd.d

aaa.a

=

lower limit of channel 1

bbb.b

=

upper limit of channel 1

ccc.c

=

lower limit of channel 2

ddd.d

=

upper limit of channel 2

For example, water activity is scaled to 0...1.00 on channel 1 and temperature
to -40...+100

°

C on channel 2:

>ASCL 0 1 -40 100
Ch1

(aw) lo

0.000

Ch1

(aw) hi

1.000

Ch2

(T ) lo

-40.000 'C

Ch2

(T ) hi

100.000 'C