beautypg.com

Vaisala HMP228 User Manual

Page 65

background image

HMP228 Transmitters

M210282en-A

Appendix 1: Serial commands

59

When asking the current address, no address number is given:

>ADDR
Address : 2 ?

RESET

Resetting the transmitter

RESET

Resets the transmitter. All settings that have been changed stay in the memory
even after reset or power failure.

Operation modes

SMODE

Setting the serial interface

SMODE xxxx

xxxx

=

STOP, RUN or POLL

In STOP mode: measurements output only by command, all commands can be

used

In RUN mode:

outputting automatically, only command S can be used

In POLL mode: measurements output only with command SEND. When in

POLL mode, the output state is changed as follows:

OPEN aa

SMODE xxxx

aa

=

address of the transmitter

xxxx

=

STOP, RUN or POLL

The OPEN command sets the bus temporarily in STOP MODE so that the
SMODE command can be given. For example:

>SMODE

which mode is in use at the moment

Serial mode : STOP

>SMODE STOP

setting STOP mode

Serial mode : STOP