4.8.1.1

Using serial commands

ACAL

The outputs on channels 1 and 2 are measured and the measured values (mA
or V) entered as calibration coefficients.

Example: both channels have 0...10 V outputs (set with AMODE command);

enter the voltages measured at the analogue outputs:

>ACAL
Ch1

U1 ( V ) ? 0.123

Ch1

U2 ( V ) ? 9.980

Ch2

U1 ( V ) ? 0.120

Ch2

U2 ( V ) ? 9.980

4.8.1.2

Using display/keypad commands

Connect an ammeter/voltmeter to the output of channel 1, select Cali
in the main menu and Analog outputs in the Cali menu. The follow-
ing is displayed (the quantity can be either mA or V):

Enter the measured lower end current/voltage on channel 1.

Enter the measured upper end current/voltage on channel 1.

Connect the meter to the output of channel 2 and enter the measured
lower end current/voltage on channel 2.

Enter the measured upper end current/voltage on channel 2.