HMP228 Transmitters
Appendix 1: Serial commands

M210282en-A

58

NOTE

The serial bus settings become effective only after reset.

When the half-duplex mode is set, it will automatically turn the echo off. Even
then the ECHO command can indicate that echo is on.

UNIT

Selecting the output units

UNIT x

x

=

m(etric units)
n(on-metric units)

metric

non-metric

aw

(aw)

(aw)

T

°

C

°

F

For example, the command for setting the non-metric units is:

>UNIT N
Output units : non metric

When the command is given with no parameters, the transmitter outputs the
currently valid setting.

ADDR

Setting the transmitter address

ADDR aa

aa

=

address (0...99)

The address is used when more than one transmitter is connected to one serial
bus. The ADDR command makes it possible to communicate with one
transmitter at a time in POLL state.

For example, transmitter is given address 99

>ADDR
Address : 2 ? 99