HMP228 Transmitters
Appendix 1: Serial commands

M210282en-A

50

The command can also be divided into two commands, so the computer can be
used for other purposes while waiting for the sensor to stabilize to the higher
humidity.

>FCRH 1
RH : 1.90

Ref1 ?

11.3

FCRH 2 ...
RH : 76.30

Ref2 ?

74.9

If the stabilization of the sensor to the humidity in the calibrator needs to be
monitored, the measurement output can be repeated by giving command c
at Ref1 and Ref2:

>FCRH
RH : 12.00

Ref1 ? c

RH : 11.70

Ref1 ? c

RH : 11.50

Ref1 ? 11.3

Press any key when ready ...
RH : 76.00

Ref2 ? 75.5

CT

Temperature calibration

Disconnect the security lock jumper!

CT

Using this command the transmitters can be calibrated against an accurate ref-
erence, such as a Pt 100 simulator. A two-point calibration is performed as
follows:

>CT
T : 0.80

Ref1 ?

0.0

Press any key when ready ...
T : 56.20

Ref2 ?

55.0

In one-point offset correction, the Ref2 prompt is acknowledged with :

>CT
T : 0.80

Ref1 ?

0.0

Press any key when ready ...
T : 75.50

Ref2 ?

If the stabilization of the sensor to the temperature of the calibrator or the ref-
erence needs to be monitored, the measurement output can be repeated by giv-
ing command c at Ref1 and Ref2:

>CT
T : 0.80

Ref1 ?

c

T : 0.40

Ref1 ?

0.00

Press any key when ready ...
T : 56.20

Ref2 ?

55.0