Chapter 7 _____________________________________________________________ Technical Data
VAISALA ________________________________________________________________________ 65
CHAPTER 7
TECHNICAL DATA
This chapter provides the technical data of the HMM100.
Specifications
Performance
Relative Humidity
Measurement range
0 ... 100 %RH
Accuracy (including non-linearity,
hysteresis, and repeatability)
Temperature range
-20 ... +40 ºC (-4 ... +104 ºF)
0…90 %RH
±2 %RH
90…100 %RH
±3 %RH
Temperature range
-40 ... -20 ºC (-40 ... -4 ºF),
+40 ... +180 ºC (+104 ... +356 ºF)
0…90 %RH
±2.5 %RH
90…100 %RH
±3.5 %RH
Factory calibration uncertainty (+20 °C) ±1.5 %RH
Humidity sensor
Vaisala HUMICAP 180R
Dewpoint Temperature
Measurement range
-20 ... +100 ºC (-4 ... +212 ºF) T
d
Accuracy (including non-linearity,
hysteresis, and repeatability) when
dewpoint depression <20 °C
(Ambient temperature - dew point)
±2 °C T
d