beautypg.com

Technical data, Specifications, Performance – Vaisala HMM100 User Manual

Page 67: Relative humidity, Dewpoint temperature, Chapter 7, Chapter 7, technical data

background image

Chapter 7 _____________________________________________________________ Technical Data

VAISALA ________________________________________________________________________ 65

CHAPTER 7

TECHNICAL DATA

This chapter provides the technical data of the HMM100.

Specifications

Performance

Relative Humidity

Measurement range

0 ... 100 %RH

Accuracy (including non-linearity,

hysteresis, and repeatability)

Temperature range

-20 ... +40 ºC (-4 ... +104 ºF)

0…90 %RH

±2 %RH

90…100 %RH

±3 %RH

Temperature range

-40 ... -20 ºC (-40 ... -4 ºF),

+40 ... +180 ºC (+104 ... +356 ºF)

0…90 %RH

±2.5 %RH

90…100 %RH

±3.5 %RH

Factory calibration uncertainty (+20 °C) ±1.5 %RH

Humidity sensor

Vaisala HUMICAP 180R

Dewpoint Temperature

Measurement range

-20 ... +100 ºC (-4 ... +212 ºF) T

d

Accuracy (including non-linearity,

hysteresis, and repeatability) when

dewpoint depression <20 °C

(Ambient temperature - dew point)

±2 °C T

d