beautypg.com

Maintenance, Changing the filter, Changing the sensor – Vaisala HMM100 User Manual

Page 53: Chapter 5, Chapter 5, maintenance

background image

Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 51

CHAPTER 5

MAINTENANCE

This chapter provides information that is needed in basic maintenance of

the HMM100.

Changing the Filter

The filter on the probe should be replaced when it is damaged or dirty.

Do not attempt to clean the filter using compressed air, solvents, or

mechanical action. Wiping the filter may block its pores and/or deposit

residue on the filter.

1.

Turn the filter counter-clockwise to loosen it.

2.

Remove the filter from the probe. Be careful not to touch the

sensors with the filter. Without the filter in place, the sensors are

easily damaged – handle the probe carefully.

3.

Check that the O-ring is undamaged and in place.

4.

Install a new filter on the probe, and tighten it so it is finger-tight.

Make sure the filter sits straight and meets the threads properly.

New filters can be ordered from Vaisala. For available filters, see section

Filter Types on page 14. The order codes of the filters are listed in

section Options and Accessories on page 68.

Changing the Sensor

The user can replace the HUMICAP180R sensor with a new one. The

new sensor must be of the HUMICAP180R type. Do not attempt to use a

different sensor type.

NOTE

Replacing the sensor is repair maintenance, and it is not necessary in

normal operation. If the accuracy of the module does not seem to be

within specification, it is more likely that the module is in need of

calibration and adjustment, and not sensor replacement. Refer to section

Calibration on page 53.