Figure 16, Changing the sensor – Vaisala HMM100 User Manual
Page 54
User's Guide _______________________________________________________________________
52 ___________________________________________________________________ M211024EN-C
1.
Remove the filter from the probe by turning it counter-clockwise.
2.
Remove the damaged sensor and insert a new one. Handle the new
sensor by the plastic socket. DO NOT TOUCH THE SENSOR
PLATE.
3.
Perform a calibration and adjustment using the CRH command as
instructed in section Humidity Calibration (CRH) on page 58.
4.
Attach a new filter on the probe. When using the stainless steel
filter, take care to tighten the filter properly (recommended force
5 Nm).
1209-033
Figure 16
Changing the Sensor
The following numbers refer to Figure 16 above:
1 = The HUMICAP 180R sensor
2 = Pt1000 sensor (do not touch)