Figure 16, Changing the sensor – Vaisala HMM100 User Manual

Page 54

background image

User's Guide _______________________________________________________________________

52 ___________________________________________________________________ M211024EN-C

1.

Remove the filter from the probe by turning it counter-clockwise.

2.

Remove the damaged sensor and insert a new one. Handle the new

sensor by the plastic socket. DO NOT TOUCH THE SENSOR

PLATE.

3.

Perform a calibration and adjustment using the CRH command as

instructed in section Humidity Calibration (CRH) on page 58.

4.

Attach a new filter on the probe. When using the stainless steel

filter, take care to tighten the filter properly (recommended force

5 Nm).

1209-033

Figure 16

Changing the Sensor

The following numbers refer to Figure 16 above:
1 = The HUMICAP 180R sensor

2 = Pt1000 sensor (do not touch)