beautypg.com

Filter types, Figure 3, Table 4 – Vaisala HMM100 User Manual

Page 16: Filter properties

background image

User's Guide _______________________________________________________________________

14 ___________________________________________________________________ M211024EN-C

Filter Types

The available filters for HMM100 are shown in Figure 3 below. All

filters are 12 mm in diameter, and can be used in both the plastic and

stainless steel probe. Note the operational temperature range of each

filter.

For order codes, see section Options and Accessories on page 68.

1204-001

Figure 3

Filter Types

Table 4

Filter Properties

No.

Filter

Diameter

Pore Size Temperature Range

1

Plastic grid filter

(fastest response time)

12 mm

-

-20 ... +80 ºC

(-4 ... +176 ºF)

2

Membrane filter

12 mm

0.2 µm

-20 ... +80 ºC

(-4 ... +176 ºF)

3

Porous PTFE filter

(general purpose)

12 mm

8 µm

-70 ... +180 ºC

(-94 ... +356 ºF)

4

Stainless steel

sintered filter

12 mm

38 µm

-70 ... +180 ºC

(-94 ... +356 ºF)

5

Catalytic VHP filter

(VHP applications)

12 mm

38 µm

-70 ... +180 ºC

(-94 ... +356 ºF)

The catalytic VPH filter is recommended for environments where

vaporized hydgoren peroxide (VHP) is used. It is mechanically similar to

the stainless steel sintered filter, but it can be identified by the line that

has been machined to its base; see Figure 3 above. For more

information, see section Catalytic VHP Filter on page 15.